President University Repository

Simulation analysis of equivalent circuit model of skin-electrode impedance for transcutaneous electrical stimulation

Show simple item record

dc.contributor.author Simatupang, Joni Welman
dc.contributor.author Wilbert Wijaya
dc.contributor.author David Tyler
dc.contributor.author and Clementine Mavridis
dc.date.accessioned 2023-04-11T02:58:37Z
dc.date.available 2023-04-11T02:58:37Z
dc.date.issued 2021
dc.identifier.issn 2302-9285
dc.identifier.uri http://repository.president.ac.id/xmlui/handle/123456789/11097
dc.description Bulletin of Electrical Engineering and Informatics Vol. 10, No. 4, August 2021, pp. 1936~1943 en_US
dc.description.abstract For more than 50 years, transcutaneous electrical stimulation method has been used to cure the spinal cord injury, stroke or cerebral palsy. This method works by activating the excitable nerves, muscle fibers by electrical current stimulation through electrode to skin interface. Electrode to skin interface requires equivalent circuit to overcome the inability of measuring the skin resistivity directly. We have learned several previous models, which are from Lawler, Moineau and Keller and Kuhn. Unfortunately, Moineau model neglects the capacitance effect, while Lawler and Keller and Kuhn include capacitive and resistive nature of skin in their equivalent circuits. Both models consisted of only one parallel RC block. Therefore, this paper presents the simulation results of the proposed equivalent circuit model using two parallel RC circuits. Simulation of the proposed model is conducted in MATLAB 2015a and compared with two previous models using certain parameters. Results show that the proposed model obtained the impedance of 10.830 kΩ when it is simulated using 100Hz frequency, for Lawler model the impedance is 5.340 kΩ and Keller and Kuhn model the impedance obtained is 6.490 kΩ. The proposed model has the refined impedance compared with other models and is expected to deliver better electrical stimulation. en_US
dc.language.iso en_US en_US
dc.publisher BEEI en_US
dc.subject Capacitive en_US
dc.subject Electrical stimulation en_US
dc.subject Impedance en_US
dc.subject Resistive en_US
dc.subject Transcutaneous en_US
dc.title Simulation analysis of equivalent circuit model of skin-electrode impedance for transcutaneous electrical stimulation en_US
dc.type Article en_US


Files in this item

This item appears in the following Collection(s)

Show simple item record

Search Repository


Advanced Search

Browse

My Account